Microwave Sample Preparation Platform System Introduction

Model:Muiltwave 3000, Anton Paar Year:2004
Location: CNST ROOM 212 CNST Map NTU Map

Theory: It is heated by microwave, the power can supply at most 1400 Watt.The limit of pressure and temperaure are 60 bar and 240 Centigrade, respectively.
Application:
1. microwave digestion
2. hydrothermal experiment
3. crystal growth
| 1. Vessel jacket made of ceramic
2. PTFE-TFM liner
3. Venting valve integrated in the seal holder and safety venting groves in the liner
4. Total volume of 100 mL
5. Filling volume between 6-50 mL
6. Patented air cooling
7. Operating pressure: 60 bar
8. Opening pressure of safety disk: 120 bar
Test pressure: 140 bar (2000 psi)
Max. Temperature: 260 Centigrade
|
Back >>
|